1130B-Series Probes User’s Guide
3
Calibrating Probes
Figure 50
Good Connection
Figure 51
Bad Connection
Step 3. Running the Probe Calibration and Deskew
1
On the Infiniium oscilloscope in the Setup menu, select the channel connected
to the probe.
2
In the Channel Setup dialog box select the Probes... button.
3
In the Probe Setup dialog box select the Calibrate Probe... button.
4
In the Probe Cal dialog box select the Calibrated Atten/Offset radio button.
5
Select the Start Atten/Offset Calibration... button and follow the on-screen
instructions for the vertical calibration procedure.