9.1 Adjustment test with external weight
Condition
:
Menu setting „E-tESt“ / Menu element 4
At verified balances, the adjustment test is locked by a switch (except accuracy class I).
To disable the access lock, destroy the seal and actuate the adjustment switch.
Fehler!
Verweisquelle konnte nicht gefunden werden.
Position of the adjustment switch see
chap.
Fehler! Verweisquelle konnte nicht gefunden werden.
Attention
:
After destruction of the seal the balance must be re-verified by an authorised agency
and a new verification wire/seal mark fitted before it can be reused for applications
subject to verification.
In weighing mode press
. E-tESt is displayed.
If „E-tESt“ is not displayed, press
to return into weighing mode and activate
menu element 4, see menu overview.
Press
, checking is started. The displayed weight value
of the recommended adjustment weight (see chapter 1) flashes.
If the weight value shall be changed, press
, the active digit flashes. Carry out
the desired setting using navigation buttons (see chapter 3.3
„Numeric input“).
Carefully place adjusting weight in the centre of the weighing
plate ,press
.
Wait until the zero display flashes.
Remove adjustment weight and press
.
Wait for a short period of time, the difference to the previous
adjustment is displayed.
Either
press
, the d-value is reset to zero.
Adjustment of balance by resetting.