Adjustment of Linearity
Go into Service menu and set the balance so that wAd and tAd are usable.
After the balance has been turned on for a sufficient amount of time at a stable room
temperature, repeat the loading/ unloading of the following weights.
The count data values are obtained by converting the absolute value of the weight used
into the number of digits required.
The values of the weights to be prepared and the count data values are stated below.
For PBJ 620-3NM
Value of Weight
Used
Count Data Value
Weight 1
100g
00100000
Weight 2
200g
00200000
Weight 3
300g
00300000
Weight 4
400g
00400000
Weight 5
500g
00500000
Weight 6
620g
00620000
For PBJ 4200-2NM
Value of Weight
Used
Count Data Value
Weight 1
500g
00050000
Weight 2
1000g
00100000
Weight 3
2000g
00200000
Weight 4
3000g
00300000
Weight 5
4000g
00400000
Weight 6
4200g
00420000
For PBJ 6200-2NM
Value of Weight
Used
Count Data Value
Weight 1
1000g
00100000
Weight 2
2000g
00200000
Weight 3
3000g
00300000
Weight 4
4000g
00400000
Weight 5
5000g
00500000
Weight 6
6200g
00620000
For PBJ 8200-1NM
Value of Weight
Used
Count Data Value
Weight 1
1000g
00100000
Weight 2
2000g
00200000
Weight 3
4000g
00400000
Weight 4
6000g
00600000
Weight 5
7000g
00700000
Weight 6
8200g
00820000