Keito K7
Technical and maintenance
Manual
INT.022EN.02
Revision “I”
24/09/2010
Page 21 of 43
Proceed as follows:
Key
Function to execute or select
Terminal Display
'Z' or 'ESC' Goes to option 0 of the main menu.
MAIN MENU:
0-SYSTEM ADJUSTMENTS
SPACE=NEXT
ENTER=SELECT
3
Select the blood pressure adjustments.
SUB-MENU PRESSURE :
3.0-BLOOD PRESSURE
SPACE=NEXT
ENTER=SELECT
1
The terminal screen will show the zero counting corresponding to
the pressure system. When it is stable press ENTER.
ZERO COUNTING: 304
ENTER
Indicate with the terminal the pressure that we are going to apply
with the manomenter to make the calibration. (ex. 200 mmHg)
INDICATE THE PRESSURE FOR
CALIBRATION AND PRESS
ENTER: 000,00
200
Now apply the same pressure with the manometer.
INDICATE THE PRESSURE FOR
CALIBRATION AND PRESS
ENTER: 200,00
ENTER
When pressing ENTER the unit calibrates the system.
During this
process the pressure in the manometer must be stable.
CALIBRATING...
At the end of the calibration the programme goes back to option 1
of the blood pressure sub-menu. In this moment, you must open
the valve of the manometer and let the cuf deflate in full.
SUB-MENU PRESSURE :
3.1-CALIBRATION
SPACE=NEXT
ENTER=SELECT
3.2 - Verify Pressure.
This option makes different continuous pressure measurements and shows them in the display. Steps to follow.
Proceed as follows:
Key
Function to execute or select
Terminal Display
'Z' or 'ESC' Goes to option 0 of the main menu.
MAIN MENU:
0-SYSTEM ADJUSTMENTS
SPACE=NEXT
ENTER=SELECT
3
Select the blood pressure adjustments.
SUB-MENU PRESSURE :
3.0-BLOOD PRESSURE
SPACE=NEXT
ENTER=SELECT
2
Option 2 measures the pressure applied with the manometer and
shows it on the display. Before doing this operation
make sure
that there is no pressure in the pneumatic ciruit
because the
unit will take the existing pressure in this moment
as zero value of
pressure
.
On the terminal screen, you will see three counters: the first and
second ones on the right show the current pressure and the one in
the left one shows the loss of pressure for every second
PRESSURE mmHg:
0 0,0 0,0
Apply some pressure with the manometer and verify that this
pressure corresponds to the one shown on the display. (ex. 200)
PRESION mmHg:
200 200,0 0,0
ENTER
Pressing any key (ENTER for example) the programme goes back
to option 1 of the pressure sub-menu.
SUB-MENU PRESSURE:
3.1-CALIBRATION
SPACE=NEXT
ENTER=SELECT