1 - Adjustments of the height system. Option 1 of the main menu.
The measurement of height is based on a system of calculation of distance by means of ultrasounds. The main components
of the system are the acoustic transducer (located at the top of the unit) and the electronic measurement circuit, which is
controlled by the Main Board P1PC. This chapter depicts the procedure to adjust the height system.
1.0 - Reset of the height system
During the autotest process, the unit makes an auto-zero of the height system. With this option of the menu, this process
can be made without having to switch the unit off and on again.
When making an auto-zero of the height system, the unit understands that there is nothing on the platform. The value of the
measured signal at this moment is taken as the zero value (0 cm) of the height system. This is why it is so
important to
check that there is nothing on the platform before doing this process
.
Proceed as follows:
Key
Function to execute or select
Terminal Display
'Z' or 'ESC' Goes to option 0 of the main menu.
MAIN MENU:
0-SYSTEM ADJUSTMENTS
SPACE=NEXT
ENTER=SELECT
1
Select the height system adjustments
SUB-MENU HEIGHT:
1.0-RESET
SPACE=NEXT
ENTER=SELECT
0
Option 0 resets (autozero) the height system.
WAIT A MOMENT...
When it is finished the program goes back to option 0 of the height
adjustment sub-menu.
SUB-MENU HEIGHT:
1.0-RESET
SPACE=NEXT
ENTER=SELECT
1.1 - Continuous height normal measurement
This action allows you to make different measurements of height with the normal frequency of measurement, and show
them in the display. This is the way that the system measures the height during normal operation.
Proceed as follows:
Key
Function to execute or select
Terminal Display
'Z' or 'ESC' Goes to option 0 of the main menu.
MAIN MENU:
0-SYSTEM ADJUSTMENTS
SPACE=NEXT
ENTER=SELECT
1
Select the height system adjustments
SUB-MENU HEIGHT:
1.0-RESET
SPACE=NEXT
ENTER=SELECT
1
The unit measures the height continuously and shows result on the
display. Ignore the other data shown in the display.
Hei.: 0cm
Amx:208cm.
Zero: 24278 Cnt:24274
NumP: 1000 Blk:21200
Tout: 32767 CodErr: 0
ENTER
When pressing any key (ENTER for example) the process ends
and the program goes back to option 1 of the height system sub-
menu.
SUB-MENU HEIGHT:
1.1- NORMAL MEASURE
SPACE=NEXT