background image

 

 

Keito K7 

Technical and maintenance 

Manual 

 

INT.022EN.02 

Revision “I” 

24/09/2010 

Page 20 of 43 

 

3 - Pressure adjustments. Option 3 of the main menu. 

3.0 - Measuring the blood pressure. 

 
This option lets you make as many blood pressure measurements as you wish. 
 

It is essential to measure the height before the blood pressure measurement. Otherwise, the 
readings of the blood preSsure will not be correct. 

 
Proceed as follows: 

Key 

Function to execute or select 

Terminal Display  

'Z' or 'ESC'  Goes to option 0 of the main menu. 

 

 

MAIN MENU: 
0-SYSTEM ADJUSTMENTS  

SPACE=NEXT 

ENTER=SELECT 

Select the blood pressure adjustments. 

SUB-MENU PRESSURE : 

3.0-BLOOD PRESSURE 
SPACE=NEXT 

ENTER=SELECT 

Blood pressure measurement. The results appear on the displays. 

SUB-MENU PRESSURE : 

3.0- BLOOD PRESSURE 
SPACE=NEXT 

ENTER=SELECT 

  

At the end of the measurement, the program goes back to option 0 
of the blood pressure adjustment sub-menu. 

SUB-MENU PRESSURE : 

3.0- BLOOD PRESSURE 

SPACE=NEXT 
ENTER=SELECT 

 
3.1 - Calibration of the blood pressure. 

 
The calibration is made in two steps: 
 

·

  The unit measures the value of the atmospheric pressure. In this moment, you must not apply any pressure to the 

sensor, and the valve of the rubber bulb must be open. 

·

  A pressure of 200 mmHg is applied to the column and the unit measures its value. 

 

Important Note:

 

The pressure adjustment is made automatically by the unit, by means of internal calculations, based on the indicated data. 
To obtain a correct calibration, take the following into account: 
  

·

 

Use a mercurial column or a manometer duly calibrated and certified. 

  

·

 

Make sure that before inflating the mercurial column or manometer, the indication of the scale is 0 mmHg. 

  

·

 

Make  sure  that  there  is  no  leakage  in  the  tubing.  It  is  also  important  (during  the  second  step,  when  applying 
pressure to the sensor) that the pressure is constant and stable during the process of adjustment. 

  

·

 

If necessary, add a supplementary reservoir to the tubing so that the capacity of the air circuit is bigger and the 
leakage is minimized. 

 

Before, you must connect the manometer, with the valve open, to the pressure sensor so that the valve and the 

compressor are out of the circuit. The resulting pneumatic system must be composed of the pressure sensor, the cuff 
(used as reservoir) and the manometer. 

Summary of Contents for K7

Page 1: ...Keito K7 Technical and maintenance Manual INT 022EN 02 Revision I 24 09 2010 Page 1 of 43 MULTIFUNCTION K7 Technical Manual...

Page 2: ...going to the computer 10 Switch the K7 on It is ready for use B Operating in the Advanced Programming Mode B 1 Description of the programming terminal The terminal has got an LCD screen with 4 lines o...

Page 3: ...in menu if you press the SPACE key the options of the current sub menu will appear If when in a sub menu you press the ENTER key the program will execute the selected function If when in a sub menu yo...

Page 4: ...Blood pressure measurement 1 Pressure calibration 2 Pressure continuous measurement 3 Pressure adjust 3 Escapement test 0 Impedance Test 1 Body fat calibration 4 Body fat and heart rate adjust 2 Heart...

Page 5: ...tarts the memory reset RESETTING THE MEMORY Once the reset is finished the program goes back to option 0 of the system adjustments sub menu SUB MENU SYSTEM 0 0 MEMORY RESET SPACE NEXT ENTER SELECT The...

Page 6: ...n if they don t belong to the pharmacy Prints the uses with smart card in the collection report Flag n 4 0 Format of the date DD MM YY Day Month Year 1 Format of the date MM DD YY Month Day Year 2 For...

Page 7: ...x confirmation on the body fat measurement Prints the graphic of the heart rate analysis horizontally 4 Always measures the body fat 5 Body fat only in full operation The user must confirm the sex on...

Page 8: ...sks once more Insert another coin 6 Once the minimum amount is reached the unit does not ask to insert more coins 7 Even if the minimum amount for use has been reached the unit asks once more Insert a...

Page 9: ...cle disabled 6 The measure of the body fat is cancelled if the user steps off the platform without indicating the sex 7 Waits for the user to indicate the sex after the fat measure does not abort the...

Page 10: ...flags The display show the uccrent value of the first flag INDICATE THE VALUE OF FLAG 1 3 AND PRESS ENTER ENTER Press ENTER until you get the flag number that you wish to modify in this example we wis...

Page 11: ...ET SPACE NEXT ENTER SELECT 3 Select the programming of the price The display shows the price for the full operation currently programmed This price is correct in our example so you press ENTER to go t...

Page 12: ...ight B press 5 Value in uros 9 0 Contabilized uros 9 4 Cummulated Weight and height 15 Weight height B press 10 Value in uros 16 0 Contabilized uros 16 4 Example of a collection ticket of a unit progr...

Page 13: ...eport and the program goes back to the system adjustments sub menu SUB MENU SYSTEM 0 5 MAINTENANCE SPACE NEXT ENTER SELECT 0 6 Coin selector test This option allows the correct detection and operation...

Page 14: ...Clear the counters The collection report has got two counters one is set to zero every time that a report is printed but the other one the cumulated keeps the total number of coins The only way to set...

Page 15: ...Z or ESC Goes to option 0 of the main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENTER SELECT 1 Select the height system adjustments SUB MENU HEIGHT 1 0 RESET SPACE NEXT ENTER SELECT 0 Option 0 r...

Page 16: ...mory when making a memory reset This adjustment will only be necessary if for any reason the inner height of the unit needs to be changed During the autotest process the unit makes an autozero of the...

Page 17: ...the frequency is modified Let us suppose that we want to set the frequency to 52 29 kHz Proceed as follows Key Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main menu...

Page 18: ...SC Goes to option 0 of the main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENTER SELECT 2 Select the weight system adjustments SUB MENU WEIGHT 2 0 RESET SPACE NEXT ENTER SELECT 0 Option 0 resets a...

Page 19: ...tc since they can cause wrong measurements to the system Proceed as follows Key Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main menu Make sure that there isn t any...

Page 20: ...steps The unit measures the value of the atmospheric pressure In this moment you must not apply any pressure to the sensor and the valve of the rubber bulb must be open A pressure of 200 mmHg is appl...

Page 21: ...ometer and let the cuf deflate in full SUB MENU PRESSURE 3 1 CALIBRATION SPACE NEXT ENTER SELECT 3 2 Verify Pressure This option makes different continuous pressure measurements and shows them in the...

Page 22: ...the decay rate 3 When the pressure is 50mmHg a ticket is printed with the elapsed time since the pressure was 250mmHg until reaching 50mmHg and with the maximum and minimum decay rate registered 4 Dur...

Page 23: ...he main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENTER SELECT 4 Select the body fat adjustments SUB MENU FAT PULSE 4 0 IMPEDANCE TEST SPACE NEXT ENTER SELECT 0 This option measures the impedance...

Page 24: ...mation In this case the electrodes record the electrical activity of the heart and from these signals the heart rate is measured At the end of the test the unit prints a ticket with the value of the p...

Page 25: ...Keito K7 Technical and maintenance Manual INT 022EN 02 Revision I 24 09 2010 Page 25 of 43 Heart rate analysis Heart rate 86 1 P mn maximum Rate 90 P mn minimum Rate 83 P mn...

Page 26: ...tial autotest and detects any malfunction Proceed as follows Key Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENT...

Page 27: ...ADJUSTMENTS SPACE NEXT ENTER SELECT 5 Select the printer adjustments SUB MENU PRINTER 5 0 RESET SPACE NEXT ENTER SELECT 1 Select the paper feed Make sure that the paper comes up centered and straight...

Page 28: ...the programme goes back to option 3 of the printer adjustment sub menu SUBMENU PRINTER 5 3 TEST Num 2 SPACE NEXT ENTER SELECT 5 4 Printer test n 3 This option starts the printer and prints simulated...

Page 29: ...n 5 of the printer adjustment sub menu SUB MENU PRINTER 5 5 CUTTER TEST SPACE NEXT ENTER SELECT 5 6 Temperature of the printer This test shows the status of the printer head without printing Proceed a...

Page 30: ...PACE NEXT ENTER SELECT 6 1 Programming the time Suppose that you wish to program 13 00 00 h Proceed as follows Key Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main...

Page 31: ...he memory SUB MENU DATE TIME 6 2 DATE SPACE NEXT ENTER SELECT Note If during the time programming you need to cancel the process press the ESC key The keys and let you move the cursor through the disp...

Page 32: ...left Moves the cursor to the right If you wish a double wide character press this key before the desired character CRTL C This combination does the same function as ENTER but it also centers the text...

Page 33: ...if a card belongs to that premise or group of premises or not erase the data contained in the card charge new credits and list the recorded measurements and or the identification data of the card If...

Page 34: ...the program goes back to option 1 of the smart card sub menu SMART CARD 8 1 CARD RESET SPACE NEXT ENTER SELECT 8 2 Programming the credit in the smart card This option is very similar to the previous...

Page 35: ...ELECT 9 Voice Buzzer Test Option 9 of the main menu With this option we can check the correct operation of the audio system of the unit as well for the buzzer as for the voice messages The volume of t...

Page 36: ...Function to execute or select Terminal Display Z or ESC Goes to option 0 of the main menu MAIN MENU 0 SYSTEM ADJUSTMENTS SPACE NEXT ENTER SELECT 9 Select the voice and buzzer test SUB MENU VOICE BUZZ...

Page 37: ...g is allright the menu goes to the previous screen In case of error the screen would be the one beside MODEM NOT RESPONDING PRESS ANY KEY TO END 10 1 Reset step by step This option is similar to the p...

Page 38: ...case of error the screen would be the one beside MODEM NOT RESPONDING PRESS ANY KEY TO END 10 3 Central telephone With this option you indicate to the unit which is the telephone number of the Data Ce...

Page 39: ...er of SMS to be deleted ERASE SMS INDICATE SMS NUMBER 01 ENTER If the SMS has been deleted correctly the message beside appears Press any key to continue or ENTER to end MESSAGE 1 DELETED Any Key The...

Page 40: ...SMS SENDING SMS If the SMS has been sent succesfully the following message will appear SMS SENT TO DATA CENTER Any Key By pressing any key the system goes back to the selected option SUBMENU GSM SMS A...

Page 41: ...Parameter 1 Parameter 1 indicates if the unit can send an SMS when switched on or when an error is detected Value Description Description 0 Does NOT send an SMS when unit is switched on 1 Sends SMS wh...

Page 42: ...tion 0 Will never send an SMS for this reason 200 Sends an SMS with all data of status and collection and clears the collection counters when collection is 200 your currency etc Parameter 6 The unit w...

Page 43: ...essary to inflate over 300 mmHg to measure the blood pressure 2 It was not possible to inflate the cuff Possible puncture in the cuff 3 Deflated up to 60mmHg without detecting impulsions 4 Before havi...

Reviews: