Keito K7
Technical and maintenance
Manual
INT.022EN.02
Revision “I”
24/09/2010
Page 20 of 43
3 - Pressure adjustments. Option 3 of the main menu.
3.0 - Measuring the blood pressure.
This option lets you make as many blood pressure measurements as you wish.
It is essential to measure the height before the blood pressure measurement. Otherwise, the
readings of the blood preSsure will not be correct.
Proceed as follows:
Key
Function to execute or select
Terminal Display
'Z' or 'ESC' Goes to option 0 of the main menu.
MAIN MENU:
0-SYSTEM ADJUSTMENTS
SPACE=NEXT
ENTER=SELECT
3
Select the blood pressure adjustments.
SUB-MENU PRESSURE :
3.0-BLOOD PRESSURE
SPACE=NEXT
ENTER=SELECT
0
Blood pressure measurement. The results appear on the displays.
SUB-MENU PRESSURE :
3.0- BLOOD PRESSURE
SPACE=NEXT
ENTER=SELECT
At the end of the measurement, the program goes back to option 0
of the blood pressure adjustment sub-menu.
SUB-MENU PRESSURE :
3.0- BLOOD PRESSURE
SPACE=NEXT
ENTER=SELECT
3.1 - Calibration of the blood pressure.
The calibration is made in two steps:
·
The unit measures the value of the atmospheric pressure. In this moment, you must not apply any pressure to the
sensor, and the valve of the rubber bulb must be open.
·
A pressure of 200 mmHg is applied to the column and the unit measures its value.
Important Note:
The pressure adjustment is made automatically by the unit, by means of internal calculations, based on the indicated data.
To obtain a correct calibration, take the following into account:
·
Use a mercurial column or a manometer duly calibrated and certified.
·
Make sure that before inflating the mercurial column or manometer, the indication of the scale is 0 mmHg.
·
Make sure that there is no leakage in the tubing. It is also important (during the second step, when applying
pressure to the sensor) that the pressure is constant and stable during the process of adjustment.
·
If necessary, add a supplementary reservoir to the tubing so that the capacity of the air circuit is bigger and the
leakage is minimized.
Before, you must connect the manometer, with the valve open, to the pressure sensor so that the valve and the
compressor are out of the circuit. The resulting pneumatic system must be composed of the pressure sensor, the cuff
(used as reservoir) and the manometer.