Model 6487 Reference Manual
Calibration
16-13
Figure 16-3
Connections for voltage source calibration
Entering calibration dates and saving calibration
klqb
For temporary calibration without saving new calibration constants, proceed to
“Locking out calibration”.
1.
Press MENU, select CAL, then press ENTER to access the calibration menu.
2.
Use either RANGE key to display the following:
CAL: SAVE
3.
Press ENTER. The unit will prompt for today’s calibration date:
DATE: 11/15/02
4.
Use the arrow and RANGE keys to set the date, then press ENTER. The unit will
then prompt for the calibration due date:
NDUE: 11/15/03
Table 16-7
Voltage source calibration summary
Model 6487 Range
Calibration Voltages*
10V
-10.00000V
+0.00000V
+10.0000V
50V
-50.00000V
+0.00000V
+50.00000V
500V
-500.0000V
+0.00000V
+500.0000V
* Nominal values. Adjust display to agree with DMM reading.
NEXT
PREV
POWER
DISPLAY
2001
MULTIMETER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
INFO
LOCAL
EXIT
ENTER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN
SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
3
50V
PEAK
1100V
PEAK
AUTO
Model 2001 DMM
120
LINE RATING
50, 60Hz
50 VA MAX
INPUT
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
IEEE-488
CAT
I
TRIGGER LINK
RS-232
MADE IN
U.S.A.
ANALOG OUT
DIGITAL I/O
505V
MAX
V-SOURCE OUTPUT
505V PK TO CHASSIS
LO
HI INTERLOCK
!
!
FUSE
LINE
400mAT
(SB)
100 VAC
120 VAC
200mAT
(SB)
220 VAC
240 VAC
505V PK
505V PK
505V PK
Model 64
8
7
I
N
PUT
V-
S
OURCE
OUTPUT