Model 6487 Reference Manual
SCPI Reference Tables
14-9
[:DAMPing]
:OHMS
Path to control ohms:
[:STATe] <b>
Enable or disable ohms for SENSe1 data.
OFF
[:STATe]?
Query ohms SENSe1 data state.
:AVOLtage
Path to A-V ohms commands:
[:ARM]
Arm A-V ohms mode.
[:ARM]?
Query if A-V ohms is armed.
(1 = armed).
:ABORt
Abort A-V ohms mode.
:VOLTage <NRf>
Set high voltage value (-505 to 505V).
10V
:VOLTage?
Query high voltage value.
:TIME <NRf>
Set time interval for each phase.
15s*
:TIME?
Query time interval for each phase.
:POINts?
Query number of points.
:ONEShot <b>
Enable or disable one-shot mode.
ON
:ONEShot?
Query state of one-shot mode.
:CYCLes <NRf>
Set number of A-V cycles (1 to 9999).
3
:CYCLes?
Query number of A-V cycles.
:UNITs <name>
Select AMPS or OHMS units.
AMPS
:UNITs?
Query units.
:CLEar
Clear A-V ohms buffer.
:AUTO <b>
Enable/disable A-V buffer auto clear.
ON
:AUTO?
Query auto clear state.
:BCOunt?
Query number of A-V cycles that have
been completed and are averaged to make
up present buffer.
*15s for 1 PLC or greater, 1s for 0.1 PLC, and 0.1s for 0.02 PLC.
Table 14-4 (cont.)
SENSe command summary
Command
Description
Default
parameter
Ref.
Section SCPI