Specifications A-11
Frequency Counter
Frequency/Period Input Characteristics and Accuracy
90 Days, 1 Year, or 2 Years
Frequency
Period
Minimum Signal Level
2
Maximum
Trigger
Accuracy
Range
1
Range
Resolution
1Hz–1MHz
1–5MHz 5–15MHz
Input
Level
±
(% of reading)
AC Voltage Input
1Hz–15 MHz
67 ns – 1 s
5 digits
60 mV
60 mV
350 mV
1100 V pk
1
0–600V
0.03
AC Current Input
1Hz– 1 MHz
1
µ
s – 1 s
5 digits
150
µ
A
1 A pk
0–600mA
0.03
Time Base
7.68MHz
±
0.01%, 0
°
C to 55
°
C.
Reading Time
420ms maximum.
Voltage Input Impedance
1M
Ω
±
2% with <140pF.
Trigger Level Adjustment
Trigger level is adjustable in 0.5% of range steps to
±
60% of
range in real-time using the up and down range buttons.
Frequency Ranging
Autoranging from Hz to MHz.
Frequency Coupling
AC + DC or AC only.
Frequency Notes
1 Subject to 2
×
10
7
V·Hz product (for inputs above 20V).
2 Valid for the lowest range. For each range increase,
multiply these numbers by 10.
Temperature (RTD)
Temperature Notes
4-Wire Accuracy
5
Range
Resolution
24 Hours
2
90 Days
3
1 Year
3
2 Years
3
–100
°
to +100
°
C
0.001
°
C
±
0.016
°
C
±
0.020
°
C
±
0.021
°
C
±
0.022
°
C
–200
°
to +630
°
C
0.001
°
C
±
0.061
°
C
±
0.066
°
C
±
0.068
°
C
±
0.070
°
C
–148
°
to +212
°
F
0.001
°
F
±
0.029
°
F
±
0.036
°
F
±
0.038
°
F
±
0.040
°
F
–328
°
to +1166
°
F
0.001
°
F
±
0.110
°
F
±
0.119
°
F
±
0.122
°
F
±
0.126
°
F
RTD Type
100
Ω
platinum, DIN 43760, 4-wire. ITS-90 (PT100, D100, F100)
and IPTS-68 (PT385, PT3916).
Sensor Current
960
µ
A (pulsed).
Temperature Coefficient
±
0.001
°
C/
°
C or
±
0.002
°
F/
°
C outside T
CAL
±
5
°
C.
Maximum Source
HI Lead Resistance
200
Ω
.
Maximum Source
LO Lead Resistance
100
Ω
.
RTD Temperature Reading Rates
1
(2- or 4-Wire)
Readings or Readings with
Time Stamp/Second
to Memory or IEEE-488
PLC
Autozero Off
Autozero On
10
3
(2.5)
1 (0.8)
2
12
(10)
4 (3.3)
1
20
(16)
17
(13)
0.1
51
(49)
41
(39)
0.01
58
(58)
46
(46)
Temperature (Thermocouple)
Thermo-
couple
Type
Range
Resolution
Accuracy
4
J
–200
°
to + 760
°
C
0.001
°
C
±
0.5
°
C
K
–200
°
to +1372
°
C
0.001
°
C
±
0.5
°
C
T
–200
°
to + 400
°
C
0.001
°
C
±
0.5
°
C
E
–200
°
to +1000
°
C
0.001
°
C
±
0.6
°
C
R
0
°
to +1768
°
C
0.001
°
C
±
3
°
C
S
0
°
to +1768
°
C
0.001
°
C
±
3
°
C
B
+350
°
to +1820
°
C
0.001
°
C
±
5
°
C
TC Temperature Reading Rates
1
Readings/Second
Readings/Second
Readings/Second
with Time Stamp
to Memory
to IEEE-488
6
to IEEE-488
6
Autozero
Autozero
Autozero
PLC
Off
On
Off
On
Off
On
10
6
(5)
2 (1.7)
6
(5)
2 (1.6)
6
(5)
2 (1.6)
2
29
(25)
9 (7.6)
29 (24)
9 (7.4)
27 (22)
9 (7.4)
1
57
(48)
47 (40)
56 (46)
46 (38)
50 (41)
42 (34)
0.1
131 (126) 107 (103)
100 (96)
84 (81)
83 (80)
72 (69)
0.01
168 (168) 112 (112)
121 (121)
89 (89)
96 (96)
74 (74)
4 Relative to external 0
°
C reference junction; exclusive of
thermocouple errors. Junction temperature may be
external. Applies for 90 days, 1 year or 2 years, T
CAL
±
5
°
C.
5 Specifications are for 10 power line cycles, autozero on,
10 reading repeat digital filter, 4-wire mode. Exclusive of
RTD probe errors.
6 Using Internal Buffer.
1 For on-scale readings, no trigger delays, digital filter off,
display off, normal autozero, internal trigger, SREAL
format. These rates are for 60Hz and (50Hz). Rates for
400Hz equal those for 50Hz. Typical values.
2 For T
CAL
±
1
°
C, following 4-hour warm-up.
3 For T
CAL
±
5
°
C, following 4-hour warm-up.
Summary of Contents for 2002
Page 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 9: ...iv...
Page 11: ...vi...
Page 55: ...Calibration 2 26...
Page 70: ...Specifications A 1 A Specifications...
Page 83: ...A 14 Specifications...
Page 101: ...Calibration Programs B 18...
Page 113: ......