Calibration Programs
B-13
'
CmdList:
DATA "Connect low-thermal short to inputs, wait 3 minutes.","DC:ZERO"
DATA "OUT 2 V","DC:V2 2"
DATA "OUT 20 V","DC:V20 20"
DATA "OUT 1 MOHM","DC:OHM1M"
DATA "OUT 100 KOHM","DC:OHM200K"
DATA "OUT 19 KOHM","DC:OHM20K"
DATA "OUT 1.9 KOHM","DC:OHM2K"
DATA "OUT 190 OHM","DC:OHM200"
DATA "OUT 19 OHM","DC:OHM20"
DATA "OUT 200 UA","DC:A200U 200E-6"
DATA "OUT 2 MA","DC:A2M 2E-3"
DATA "OUT 20 MA","DC:A20M 20E-3"
DATA "OUT 200 MA","DC:A200M 200E-3"
DATA "OUT 1A","DC:A2 1"
DATA "Disconnect calibrator from INPUT and SENSE jacks.","DC:OPEN"
DATA "Performing AC calibration, please wait...","UNPR:ACC"
DATA "OUT 20 V,1 KHZ","LLEV:STEP 1"
DATA "OUT 20 V,30 KHZ","LLEV:STEP 2"
DATA "OUT 200 V,1 KHZ","LLEV:STEP 3"
DATA "OUT 200 V,30 KHZ","LLEV:STEP 4"
DATA "OUT 1.5 V,1KHZ","LLEV:STEP 5"
DATA "OUT 0.2 V,1 KHZ","LLEV:STEP 6"
DATA "OUT 5 MV,100KHZ","LLEV:STEP 7"
DATA "OUT 0.5 MV,1 KHZ","LLEV:STEP 8"
DATA "OUT 100 V,0 HZ","LLEV:STEP 9"
DATA "OUT -20 V,0 HZ","LLEV:STEP 10"
DATA "Connect low-thermal short to rear panel INPUT jacks.","LLEV:STEP 11"
DATA "OUT 20 MA,1 KHZ","LLEV:STEP 12"
DATA "Apply 2V rms @ 1Hz from synthesizer to INPUT jacks","LLEV:STEP 13"
Summary of Contents for 2002
Page 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 9: ...iv...
Page 11: ...vi...
Page 55: ...Calibration 2 26...
Page 70: ...Specifications A 1 A Specifications...
Page 83: ...A 14 Specifications...
Page 101: ...Calibration Programs B 18...
Page 113: ......