Specifications A-5
AC Volts
ACV Input Characteristics
Temperature Coefficient
2
rms
Peak
Full Scale
±
(% of r % of range) /
°
C
Range
Input
rms
Resolution
Input Impedance
Outside T
CAL
±
5
°
C
200 mV
1 V
210.0000
100 nV
1M
Ω
±
2% with <140pF
0.004 + 0.001
2
V
8 V
2.100000
1
µ
V
1M
Ω
±
2% with <140pF
0.004 + 0.001
20
V
100 V
21.00000
10
µ
V
1M
Ω
±
2% with <140pF
0.006 + 0.001
200
V
800 V
210.0000
100
µ
V
1M
Ω
±
2% with <140pF
0.006 + 0.001
750
V
1100 V
775.000
1 mV
1M
Ω
±
2% with <140pF
0.012 + 0.001
AC Voltage Uncertainty
=
±
[ (% of reading)
×
(measured value) + (% of range )
×
(range used) ] / 100.
PPM Accuracy
= (% accuracy)
×
10,000.
0.015% of Range
= 30 counts for ranges up to 200V and 113 counts on 750V range at 5
½
digits.
Low Frequency Mode rms
1
90 Days, 1 Year or 2 Years,
±
2
°
C from last AC self-cal, for 1% to 100% of range
3
,
±
(% of r % of range)
Range
1–10Hz
5
10–50Hz
50–100Hz
0.1–2kHz
2–10kHz
10–30kHz
30–50kHz
50–100kHz 100–200kHz 0.2–1MHz 1–2MHz
200 mV
0.09+0.015
0.06+0.015
0.035+0.015
0.03+0.01
0.02+0.01
0.025+0.01
0.05+0.01
0.3+0.015
0.75+0.025
2+0.1
5+0.2
2
V
0.09+0.015
0.04+0.015
0.025+0.015
0.02+0.01
0.02+0.01
0.025+0.01
0.05+0.01
0.3+0.015
0.75+0.025
2+0.1
5+0.2
20
V
0.1 +0.015
0.06+0.015
0.035+0.015
0.03+0.015
0.04+0.015
0.05 +0.015
0.07+0.015
0.3+0.015
0.75+0.025
4+0.2
7+0.2
5
200
V
4
0.1 +0.015
0.05+0.015
0.03 +0.015
0.03+0.015
0.04+0.015
0.05 +0.015
0.07+0.015
0.3+0.015
0.75+0.025
5
4+0.2
5
750
V
4
0.13+0.015
0.09+0.015
0.05 +0.015
0.05+0.015
0.06+0.015
0.08 +0.015
0.1 +0.015
5
0.5+0.015
5
Normal Mode rms
1
90 Days, 1 Year or 2 Years,
±
2
°
C from last AC self-cal, for 1% to 100% of range
3
,
±
(% of r % of range)
Range
20–50Hz
50–100Hz
0.1–2kHz
2–10kHz
10–30kHz
30–50kHz
50–100kHz 100–200kHz 0.2–1MHz 1–2MHz
200 mV
0.25+0.015
0.07+0.015
0.02+0.01
0.02+0.01
0.025+0.01
0.05+0.01
0.3+0.01
0.75+0.025
2+0.1
5+0.2
2
V
0.25+0.015
0.07+0.015
0.02+0.01
0.02+0.01
0.025+0.01
0.05+0.01
0.3+0.01
0.75+0.025
2+0.1
5+0.2
20
V
0.25+0.015
0.07+0.015
0.03+0.015
0.04+0.015
0.05 +0.015
0.07+0.015
0.3+0.015
0.75+0.025
4+0.2
7+0.2
5
200
V
4
0.25+0.015
0.07+0.015
0.03+0.015
0.04+0.015
0.05 +0.015
0.07+0.015
0.3+0.015
0.75+0.025
5
4+0.2
5
750
V
4
0.25+0.015
0.1 +0.015
0.05+0.015
0.06+0.015
0.08 +0.015
0.1 +0.015
5
0.5+0.015
5
AC magnitude: rms or Average. Peak and Crest Factor measurements also available.
dB Accuracy rms
±
dB, 90 Days, 1 Year or 2 Years, T
CAL
±
5
°
C, Reference=1V, Autoranging, Low Frequency Mode, AC+DC Coupling
Input
1–100Hz
0.1–30kHz
30–100kHz
100–200kHz
0.2–1MHz
1–2MHz
–54 to–40 dB
(2 mV to 10mV)
0.230
0.225
0.236
0.355
–40 to–34 dB
(10 mV to 20mV)
0.036
0.031
0.041
0.088
–34 to
6
dB
(20 mV to
2
V)
0.023
0.018
0.028
0.066
0.265
0.630
6 to 26
dB
(2
V to 20
V)
0.024
0.024
0.028
0.066
0.538
0.820
5
26 to 46
dB
(20
V to 200
V)
0.024
0.024
0.028
0.066
5
0.538
5
46 to 57.8 dB (200
V to 775
V)
0.018
0.021
0.049
5
ACV Reading Rates
5,6
Readings/Second with
Measurement
Default
Readings/Second to Memory
Readings/Second to IEEE-488
12
Time Stamp to IEEE-488
12
PLC
Aperture
Bits
Digits
Autozero Off
Autozero On
Autozero Off
Autozero On
Autozero Off Autozero On
10
167 ms (200 ms)
29
6
½
6
(5)
2 (1.7)
6
(5)
2 (1.6)
6
(5)
2 (1.6)
2
33.4 ms (40 ms)
27
5
½
29
(25)
9 (7.6)
28
(23)
9 (7.4)
26
(21)
9 (7.4)
1
16.7 ms (20 ms)
26
5
½
56
(48)
47
(40)
52
(43)
44
(36)
48
(39)
40
(33)
0.2
3.34 ms
(4 ms)
23
5
½
145 (129)
110
(98)
131 (117)
100
(88)
102
(91)
79
(70)
0.1
1.67 ms
(2 ms)
22
5
½
150 (144)
112 (108)
132 (127)
101
(97)
102
(98)
80
(77)
0.02
334
µ
s (400
µ
s)
20
5
½
150 (150)
115 (115)
132 (132)
103 (103)
102 (102)
80
(80)
0.01
167
µ
s (167
µ
s)
19
4
½
382 (382)
116 (116)
251 (251)
103 (103)
163 (163)
80
(80)
0.01
8
167
µ
s (167
µ
s)
19
4
½
2000 (2000)
2000 (2000)
Summary of Contents for 2002
Page 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 9: ...iv...
Page 11: ...vi...
Page 55: ...Calibration 2 26...
Page 70: ...Specifications A 1 A Specifications...
Page 83: ...A 14 Specifications...
Page 101: ...Calibration Programs B 18...
Page 113: ......