A-8 Specifications
Ohms (cont’d)
2-Wire Resistance Reading Rates
10,12
Readings/Second with
Measurement
Default
Readings/Second to Memory
Readings/Second to IEEE-488
16
Time Stamp to IEEE-488
16
PLC
Aperture
Bits
Digits
Autozero Off
Autozero On
Autozero Off
Autozero On
Autozero Off Autozero On
10
167 ms (200 ms)
29
8
½
6
( 5 )
2 (1.7)
6
(5)
2 (1.6)
6
(5)
2 (1.6)
2
33.4 ms (40 ms)
27
7
½
29
(25)
9 (7.6)
29
(24)
9 (7.4)
27
(22)
9 (7.4)
1
16.7 ms (20 ms)
26
7
½
56
(48)
47 (40)
55
(45)
46 (38)
50
(41)
42
(34)
0.2
11
3.34 ms
(4 ms)
23
6
½
222 (197)
156 (139)
220 (196)
148 (132)
156 (139)
107
(95)
0.1
11
1.67 ms
(2 ms)
22
6
½
330 (317)
176 (169)
305 (293)
166 (159)
157 (151)
110 (106)
0.02
11
334
µ
s (400
µ
s)
20
5
½
330 (330)
182 (182)
305 (305)
172 (172)
160 (160)
113 (113)
0.01
11
167
µ
s (167
µ
s)
19
4
½
384 (384)
186 (186)
352 (352)
172 (172)
179 (179)
123 (123)
0.01
8,11
167
µ
s (167
µ
s)
19
4
½
2000 (2000)
2000 (2000)
4-Wire Resistance Reading Rates
10,12
Readings or Readings with Time Stamp/Second to Memory or IEEE-488
16
Measurement
Default
Autozero Off
Autozero Off
Autozero On
Autozero On
PLC
Aperture
Bits
Digits
Offset Comp. Off
Offset Comp. On
Offset Comp. Off
Offset Comp. On
10
167 ms (200 ms)
29
8
½
6
( 5 )
3
(2.5)
2 (1.6)
1 (0.8)
2
33.4 ms (40 ms)
27
7
½
27
(22)
13 (10.7)
9 (7.4)
4 (3.5)
1
16.7 ms (20 ms)
26
7
½
50
(41)
25
(20)
42
(34)
20 (16)
0.2
11
3.34 ms
(4 ms)
23
6
½
154
(137)
76
(68)
115 (102)
54 (48)
0.1
11
1.67 ms
(2 ms)
22
6
½
184
(176)
92
(88)
123 (118)
63 (60)
0.02
11
334
µ
s (400
µ
s)
20
5
½
186
(186)
107
(107)
126 (126)
72 (72)
0.01
11
167
µ
s (167
µ
s)
19
4
½
211
(211)
107
(107)
133 (133)
72 (72)
Ohms Notes
1 Current source has an absolute accuracy of
±
5%.
2 Refers to source lead resistance. Sense lead resistance is
limited only by noise considerations. For best results, it is
suggested that it be limited to 1.5k
Ω
.
3 Offset compensation voltage plus source current times
measured resistance must be less than source current
times resistance range selected.
4 For 2-wire mode.
5 Specifications are for 10 power line cycles, 10-reading
repeat digital filter, synchronous autozero, autorange off,
4-wire mode, offset compensation on (for 20
Ω
to 20k
Ω
ranges), except as noted.
6 For T
CAL
±
1
°
C, following 4-hour warm-up. T
CAL
is ambient
temperature at calibration (23
°
C at the factory).
7 For T
CAL
±
5
°
C, following 4-hour warm-up.
8 In burst mode, display off. Burst mode requires autozero
refresh (by changing resolution or measurement function)
once every 24 hours.
9 For T
CAL
±
5
°
C, normal autozero. 1-year and 2-year accuracy
can be found by applying the same speed accuracy ppm
changes to the 1-year or 2-year base accuracy.
10 For on-scale readings, no trigger delays, digital filter off,
internal trigger, normal autozero, display off, SREAL
format. These rates are for 60Hz and (50Hz). Rates for
400Hz equal those for 50Hz.
11 Ohms measurements at rates lower than 1 power line
cycle are subject to potential noise pickup. Care must be
taken to provide adequate shielding.
12 Typical values. Peak-to-peak noise equals 6 times rms
noise.
13 Current source is paralleled with a 10M
Ω
resistance.
14 Specifications apply for 20-reading repeat digital filter,
T
REF
±
0.5
°
C (T
REF
is the initial ambient temperature), and
for measurements within 10% of the initial measurement
value and within 10 minutes of the initial measurement
time.
15 Specifications are for 1 power line cycle, normal autozero,
digital filter off, autorange off, 4-wire mode, offset
compensation off, except as noted.
16 Using Internal Buffer.
DC Amps
DCI Input Characteristics and Accuracy
Maximum
Relative Accuracy
1
Temperature Coefficient
1
Full
Burden
±
(ppm of r ppm of range)
±
(ppm of r ppm of range)/
°
C
Range
Scale
Resolution
Voltage
6
24 Hours
2
90 Days
3
1 Year
3
2 Years
3
Outside T
CAL
±
5
°
C
200
µ
A
210.00000
10 pA
0.25 V
50 + 6
275 + 25
350 + 25
500 + 25
50 + 5
2 mA
2.1000000
100 pA
0.3 V
50 + 5
275 + 20
350 + 20
500 + 20
50 + 5
20 mA
21.000000
1 nA
0.35 V
50 + 5
275 + 20
350 + 20
500 + 20
50 + 5
200 mA
210.00000
10 nA
0.35 V
75 + 5
300 + 20
375 + 20
525 + 20
50 + 5
2
A
2.1000000
100 nA
1.1 V
350 + 5
600 + 20
750 + 20
1000 + 20
50 + 5
DC Current Uncertainty
=
±
[ (ppm reading)
×
(measured value) + (ppm of range)
×
(range used) ] / 1,000,000.
% Accuracy
= (ppm accuracy) / 10,000.
5ppm of Range
= 10 counts at 6
½
digits.
Settling Characteristics
Pre-programmed settling delay times are for <500pF external circuit capacitance. Reading
settling times are affected by source impedance and cable dielectric absorption characteristics.
Ohms Voltage Drop
Measurement
Available as a multiple display.
Autoranging
Autoranges up at 105% of range, down at 10% of range.
Summary of Contents for 2002
Page 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 9: ...iv...
Page 11: ...vi...
Page 55: ...Calibration 2 26...
Page 70: ...Specifications A 1 A Specifications...
Page 83: ...A 14 Specifications...
Page 101: ...Calibration Programs B 18...
Page 113: ......