Calibration
2-7
Step 5: Ohms calibration
1. After completing 20VDC calibration, the instrument
will display the following:
CONNECT 1 M
Ω
4W
2. Set the calibrator output to 1.00000M
Ω
, and make sure
that external sense is turned on.
NOTE
Use external sensing (4-wire ohms) when
calibrating all resistance ranges. Be sure
that the calibrator external sense mode is
on.
3. Press ENTER, and note that the Model 2002 displays
the resistance calibration value:
INPUT = 1.0000000 M
Ω
4. Using the cursor and range keys, set the resistance value
displayed by the Model 2002 to the exact resistance val-
ue displayed by the calibrator. (The allowable range is
from 475k
Ω
to 1.025M
Ω
.)
5. Press ENTER, and note that the instrument displays the
following during 1M
Ω
calibration:
Performing 1 M
Ω
Calibration
6. After completing 1M
Ω
calibration, the instrument will
display the following:
CONNECT 100 k
Ω
4W
7. Set the calibrator output to 100k
Ω
, and make sure that
external sense is turned on.
8. Press ENTER, and note that the Model 2002 displays
the resistance calibration value:
INPUT = 100.00000 k
Ω
9. Using the cursor and range keys, set the resistance value
displayed by the Model 2002 to the exact resistance val-
ue displayed by the calibrator. (The allowable range for
this parameter is from 95k
Ω
to 205k
Ω
.)
10. Press ENTER to complete the 200k
Ω
calibration step.
11. Repeat steps 7 through 10 for the 20k
Ω
, 2k
Ω
, 200
Ω
, and
20
Ω
ranges in that order. Be sure to set the calibrator to
the correct resistance value, and adjust the Model 2002
display to agree with the calibrator value.
Step 6: DC amps calibration
1. After ohms calibration is completed, the instrument will
prompt you for the first DC amps calibration step:
CONNECT 200 µADC
2. Connect the DC amps calibrator to the AMPS and
INPUT LO terminals (see Figure 2-3).
3. Set the calibrator output to 200.000µA, and make sure
the unit is in operate. (The allowable range is from 95µA
to 205µA.)
4. Be sure that the displayed current matches the calibra-
tion value, then press ENTER to complete this calibra-
tion step.
5. Repeat steps 3 and 4 for the remaining amps calibration
points as follows:
Calibration
step
Calibration
value*
Allowable range
2M
Ω
200k
Ω
20k
Ω
2k
Ω
200
Ω
20
Ω
1M
Ω
100k
Ω
19k
Ω
1.9k
Ω
190
Ω
19
Ω
475k
Ω
to 1.025M
Ω
95k
Ω
to 205k
Ω
9.5k
Ω
to 20.5k
Ω
0.95k
Ω
to 2.05k
Ω
95
Ω
to 205
Ω
9.5
Ω
to 20.5
Ω
* Nominal values shown. Use exact calibrator value.
Calibration
step
Calibration
current
Allowable range
200µA
2mA
20mA
200mA
2A
200.000µA
2.00000mA
20.0000mA
200.0000mA
1.00000A
95µA to 205µA
0.95mA to 2.05mA
9.5mA to 20.5mA
95mA to 205mA
0.95A to 2.05A
Summary of Contents for 2002
Page 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 9: ...iv...
Page 11: ...vi...
Page 55: ...Calibration 2 26...
Page 70: ...Specifications A 1 A Specifications...
Page 83: ...A 14 Specifications...
Page 101: ...Calibration Programs B 18...
Page 113: ......