Calibration Command Reference
3-8
:OHM200K
(CALibration:PROTected:DC:OHM200K)
Purpose
To program the 200k
Ω
comprehensive calibration step.
Format
:cal:prot:dc:ohm200k <Cal_resistance>
Parameter
<Cal_resistance> = 95E3 to 205E3 [
Ω
]
Description
:OHM200K programs the 200k
Ω
comprehensive calibration step. The allowable range of the
calibration resistance parameter is from 95k
Ω
to 205k
Ω
. Use a 100k
Ω
value whenever possible,
or the closest possible value.
Example
:CAL:PROT:DC:OHM200K 100E3
Program 200k
Ω
step.
:OHM20K
(CALibration:PROTected:DC:OHM20K)
Purpose
To program the 20k
Ω
comprehensive calibration step.
Format
:cal:prot:dc:ohm20k <Cal_resistance>
Parameter
<Cal_resistance> = 9.5E3 to 20.5E3 [
Ω
]
Description
:OHM20K programs the 20k
Ω
comprehensive calibration step. The allowable range of the cal-
ibration resistance parameter is from 9.5k
Ω
to 20.5k
Ω
. Use the 20k
Ω
value whenever possible,
or the closest possible value (for example, 19k
Ω
, which is the closet value available on many
calibrators).
Example
:CAL:PROT:DC:OHM20K 19E3
Program 20k
Ω
step.
:OHM2K
(CALibration:PROTected:DC:OHM2K)
Purpose
To program the 2k
Ω
comprehensive calibration step.
Format
:cal:prot:dc:ohm2k <Cal_resistance>
Parameter
<Cal_resistance> = 950 to 2.05E3 [
Ω
]
Description
:OHM2K programs the 2k
Ω
comprehensive calibration step. The allowable range of the cali-
bration resistance parameter is from 0.95k
Ω
to 2.05k
Ω
. Use the 2k
Ω
value whenever possible,
or the closest possible value (for example, 1.9k
Ω
, which is the closet value available on many
calibrators).
Example
:CAL:PROT:DC:OHM2K 1.9E3
Program 2k
Ω
step.
Summary of Contents for 2002
Page 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 9: ...iv...
Page 11: ...vi...
Page 55: ...Calibration 2 26...
Page 70: ...Specifications A 1 A Specifications...
Page 83: ...A 14 Specifications...
Page 101: ...Calibration Programs B 18...
Page 113: ......