Specifications A-13
Maximum Input Levels
Overload
Rated
Recovery
Input
1
Time
HI to LO
±
1100V
< 900 ms
HI Sense to LO
±
350V pk 250V rms
< 900 ms
LO Sense to LO
±
150V pk 100V rms
< 900 ms
I Input to LO
2A,
±
250V
(fused)
—
HI to Earth
±
1600V
< 900 ms
LO to Earth
±
500V
Note 1:
For voltages between other terminals, these ratings
can be added.
Delay and Timer
Time Stamp
Resolution:
1
µ
s.
Accuracy:
±
0.01% of elapsed time
±
1
µ
s.
Maximum:
2,100,000.000000 seconds (24 days, 7 hours).
Delay Time
(Trigger edge to reading initiation)
Maximum:
999,999.999 seconds (11 days, 14 hours).
Resolution:
1ms.
Jitter:
±
1ms.
Timer
(Reading initiation to reading initiation)
Maximum:
999,999.999 seconds (11 days, 14 hours).
Resolution:
1ms.
Jitter:
±
1ms.
IEEE-488 Bus Implementation
Implementation
IEEE-488.2, SCPI-1991.0.
Multiline Commands
DCL, LLO, SDC, GET, GTL,
UNT, UNL, SPE, SPD.
Uniline Commands
IFC, REN, EOI, SRQ, ATN.
Interface Commands
SH1, AH1, T5, TE0, L4, LE0,
SR1, RL1, PP0, DC1, DT1, C0,
E1.
Digital I/O
Connector Type
8 pin “D” subminiature.
Input
One pin, TTL compatible.
Outputs
Four pins. Open collector,
30V maximum pull-up
voltage, 100mA maximum
sink current, 10
Ω
output
impedance.
Control
Direct control by output or
set real-time with limits.
General Specifications and Standards Compliance
Power
Voltage:
90–134V and 180–250V, universal self-selecting.
Frequency:
50Hz, 60Hz, or 400Hz, self-identifying at power-
up.
Consumption:
<55VA.
Environmental
Operating Temperature:
0
°
C to 50
°
C.
Storage Temperature:
–40
°
C to 70
°
C.
Humidity:
80% R.H., 0
°
C to 35
°
C, per MIL-T-28800E
1
Para
4.5.5.1.2.
Calibration
Type:
Software. No manual adjustments required.
Sources:
2 DC voltages, 6 resistances, and 5 DC currents. All
other functions calibrated (adjusted) from these sources and
a short circuit. No AC calibrator required for adjustment.
Average Time to Perform:
40 minutes for comprehensive
calibration, 6 minutes for AC-only calibration.
Process
MIL-STD 45662A.
Physical
Case Dimensions:
90mm high
×
214mm wide
×
369mm deep
(3
½
in.
×
8
½
in.
×
14
½
in.).
Working Dimensions:
From front of case to rear including
power cord and IEEE-488 connector: 15.0 inches.
Net Weight:
<4.2kg (<9.2 lbs.).
Shipping Weight:
<9.1kg (<20 lbs.).
Standards
EMI/RFI:
Conforms to VDE 0871B (per Vfg 1046/1984), IEC
801-2. Meets FCC part 15 Class B, CISPR-22 (EN55022).
Safety:
Conforms to IEC348, CAN/CSA-C22.2. No. 231, MIL-
T-28800E
1
. Designed to UL1244.
Accessories Supplied
The unit is shipped with line cord, high performance modular
test leads, operator’s manual, option slot cover, and full
calibration data.
Note 1
For MIL-T-28800E, applies to Type III, Class 5, Style E.
Extended Memory/Non-Volatile Memory Options
DATA STORAGE
6
½
-Digit
Size
with
Setup Storage
Model
(Bytes)
4
½
-Digit
Time Stamp
Type
Number
Type
2002
8k
2,027
404
volatile
1
non-volatile
2002/MEM1
32k
6,909
1,381
non-volatile
5
non-volatile
2002/MEM2
128k
29,908
5,980
non-volatile
10
non-volatile
These are the minimum sizes to expect.
Specifications subject to change without notice.
Summary of Contents for 2002
Page 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 9: ...iv...
Page 11: ...vi...
Page 55: ...Calibration 2 26...
Page 70: ...Specifications A 1 A Specifications...
Page 83: ...A 14 Specifications...
Page 101: ...Calibration Programs B 18...
Page 113: ......