8 Safety check - Test instructions | 8.2 Ceiling mounting
Connect accessible conductive parts [ACP] to protective
earth conductor (PE)
ACP = accessible conductive parts
ACP
ACP
Note
Additional measuring points ACP X must be taken into consideration in the
presence of accessories.
ACPs on the treatment light KaVoLUX 540 LED
No ACPs need to be connected to the operating light during the measurement
with the protective conductor (PE), as all relevant parts are connected to the
protective conductor (PE) before they leave the factory and included in the test.
Measuring the protective conductor resistance
Limit
< 0,3 Ω (maximum value!)
Note
The integrity of the power supply cable, in particular the protective earth wire
of the power cable must be ensured. As this is a fixed installation, the evalu-
ation can be conducted by means of a visual inspection. If damage is determ-
ined, the further procedure to be taken is specified in the general instruc-
tions.
Note
In this measurement the resistance of the protective earth connection of the
supply network can be taken into consideration.
Note
If applicable: all removable supply connection lines, which are kept handy for
possible use, should be taken into consideration and the respective PE meas-
ured.