Instructions for use KaVoLUX 540 LED U / T
6 Operation | 6.1 Operation of the treatment light KaVoLUX 540 LED U
The KaVoLUX 540 LED operating light can be used in the following modes:
▪ Normal light: preset at 5,500 Kelvin and 30,000 Lux corresponding to day-
light quality
▪ COMPOsave mode: Enables longer working times for composites by filtering
the blue components of the light (compatible mode for light-activatable res-
toration materials).
▪ Dimmed light: approx. 4,000 Kelvin; equivalent to the light of a halogen op-
erating light
▪ Laser mode: Light mode with no detrimental effect on the KaVo KEY Laser
III, the KEY Laser 3+ or the KaVo DIAGNOdent
When operated In dimmed mode, the LED lamp functions according to a
dimmed halogen lamp. The colour temperature is approx. 4,000 K and the
composite can harden prematurely. This can have a negative impact on the
durability of the treatment.
The COMPOsave modes prevents the composite from hardening prematurely.
As opposed to the dimmed light, the blue components of the light are filtered in
the process. Therefore the composite can be processed longer in COMPOsave
mode.
Note
To ensure the safety of the patient and prevent cross infection, the magnify-
ing mirror must be used and adjusted by the office personnel.
6.1 Operation of the treatment light KaVoLUX 540 LED U
Note
The operation of the operating light KaVoLUX 540 LED via the treatment unit
is also described in the instructions for use of the unit.
6.1.1 Turning the operating light On and Off
Sensor KaVoLUX 540 LED
▶
Press the "Treatment light" button.