Instructions for use KaVoLUX 540 LED U / T
6 Operation | 5.3 Setting the light head brakes
78 / 110
6 Operation
WARNING
Unintentional activation of the KaVo KEY Laser III and KEY Laser 3+.
Simultaneous application of the operating light KaVoLUX 540 LED and the
KaVo KEY Laser III or KEY Laser 3+ can lead to the unintentional activation of
the KaVo KEY Laser III and KEY Laser 3+.
▶
When using the KaVo KEY Laser III or the KEY Laser 3+, switch the operat-
ing light to laser mode.
▶
Or switch off the operating light, do not use the KaVo KEY Laser III or
KEY Laser 3+ and the KaVoLUX 540 LED operating light simultaneously.
CAUTION
Inadvertent activation of the sensor of the KaVoLUX 540 LED operat-
ing light by laser or third-party light.
The use of a laser or third-party light can lead to malfunction or trigger the
sensor of the KaVoLUX 540 LED and switch the operating light on or off inad-
vertently.
▶
Have the sensor disabled by an authorised engineer.
WARNING
Wrong handling.
Reversible blinding (temporary sight impairment).
▶
Do not direct the light field at patients, users or/and third parties while you
switch on the light.
▶
Do not direct the light field at the patient's eyes when you move the light
head.
▶
Maintain a clearance of approx. 700 mm between the operating light and
the mouth of the patient.
CAUTION
Stroboscopic effect of the rotating instrument.
A stroboscopic effect could arise in instruments rotating at a certain speed dur-
ing application of the KaVoLUX 540 LED. This is an optical illusion that makes
the handpiece appear to be standing still or rotating extremely slowly.
Risk of injury.
▶
If the stroboscopic effect is evident, change the speed slightly and continue
working as usual.
CAUTION
Faulty measurement in connection with KaVo DIAGNOdent.
Simultaneous application of the operating light KaVoLUX 540 LED and the
KaVo DIAGNOdent can lead to faulty measurements.
▶
Switch the operating light to laser mode when using the
KaVo DIAGNOdent.
▶
Or switch off the operating light, do not use KaVo DIAGNOdent and operat-
ing light KaVoLUX 540 LED simultaneously.
CAUTION
Premature hardening of composite fillings.
A light intensity that is too high can have a negative impact on the durability of
the treatment.
▶
Select the appropriate dimming level according to the processing time.