Instructions for use KaVoLUX 540 LED U / T
4 Assembly | 4.3 Assembly of the KaVoLUX 540 LED U
36 / 110
①
CAN circuit device /
②
CAN circuit patient chair
Setting the CAN circuit device
▶
Select the device type with the serial number from the following table and
adjust the switch setting on the relevant printed circuit board.
Display
Controller
Assistant
Unit con-
trol
Dentist
control
KL Elec-
tronics
Display
Controller
Dentist
KaVoLUX
540 LED
Switch
S2
S5
S5
SW1
S2
E80 SN
100xxxxx
without 540
LED
on
off**
off
off
on
n. a.*
with 540
LED
off
off
off
off
on
fix (on)
E70 SN
200xxxxx
without 540
LED
n. a.*
on
off
off
on
n. a.*
with 540
LED
n. a.*
off
off
off
on
fix (on)
* not available / nicht vorhanden
** with Display Memospeed on the assistant element, this switch must be set
to "on"
CAN bus device
Unit control
Dentist control
KaVoLUX 540
LED
Switch
S5
E70 SN 201xxxxx
E80 SN 101xxxxx
without 540 LED
on
fix (on)
n. a.*
with 540 LED
off
fix (on)
fix (on)
* not available / nicht vorhanden
Note
To measure the CAN resistance, the unit must be switched off and the cable
connected.
▶
Switch off the units to measure the CAN resistance.