Ensure that the protective earth screw
⑥
and fastening screws
⑤
and
⑦
do
not reach into the light holder.
Note
When the mounting rod extends into the light holder, it can damage the
paint.
Screw in the depth stop screw
⑧
completely.
▶
Slide the cover ring
①
onto the mounting pole
②
.
▶
Insert the mounting pole into the lamp holder.
Positioning the mounting pole:
▪ It is approx. 1680 mm from the top edge to the floor
▪ The borehole for the swivel arm brake points in the direction 6 o´ clock
▶
Tighten the earth protector screw
⑥
and fastening screws
⑤
and
⑦
.
▶
Return the disinfection container
④
to its former position and fix with
screws
③
.
▶
Align the body of the unit so that the mounting pole is exactly vertical (see
device installation instructions).
Check with a spirit level.
▶
Push down the cover ring on the mounting pole.