Instructions for use KaVoLUX 540 LED U / T
4 Assembly | 4.3 Assembly of the KaVoLUX 540 LED U
▶
Press and hold down button "function level" to quit the service mode.
▶
Press the "Remote Control" button.
ESTETICA E70 to SN 20100000
Note
The note bit OP light LED must be set to "1" in service mode U04 Index 28 to
activate the KaVoLUX 540 LED U operating light on the treatment units
ESTETICA E70/E80.
Note
The KaVo Service Tester (
Mat. no. 10032212
) is required to make adjust-
ments in service mode.
▪ Service tester: Service tester I²C Mat. no. 1.003.2212
See also:
2
TA treatment unit: Start, continue and end the service mode
▶
To start service mode, press and hold the "Decrease value" and "Increase
value" keys.
▶
Press the foot pedal.
ð
Service mode U1 is displayed.
▶
Press the "Select level" button until "U04" is shown on the display.
▶
Press the "Increase value" button and the "Reduce value" button to set the
index to "28".
▶
Press the „Clock“ button to set the note bit to "1: OP Light LED".
▶
To quit the service mode, press and hold the "Decrease value" and "In-
crease value" keys.
▶
Press the foot pedal.
ð
This quits the service mode.