8 Safety check - Test instructions | 8.2 Ceiling mounting
Note
If no other specifications have been made, all values relating to voltage and
current are effective values of alternating voltage, direct voltage or pulsating
voltage res. alternating current, direct current or pulsating current.
Note
Connection cables such as data cables and cables for the functional earth
could simulate protective conductor connections. These types of supplement-
ary but unintentional protective earth connections could lead to erroneous
measurements.
Note
Cables and wires, e.g. power supply cords, measuring circuits and data lines,
must be arranged appropriately such that their influence on measurements is
minimised.
Note
The following measuring aids can be ordered: KaVo measuring cable (
Mat.
no. 0.411.8811
)
Using the measuring cable
①
the unit is disconnected from the mains supply
and connection of the treatment centre to the safety tester is enabled. Hence,
the customer-provided mains supply L & N on the power input board need not
be disconnected. The adapter cable
②
is included in the delivery of the KaVo
measuring cable and is required for older treatment centres that are not
equipped with an X2 connector.
97 / 110