Instructions for use KaVoLUX 540 LED U / T
10 Specifications on the electromagnetic compatibility in accordance with EN 60601-1-2 | 10.3 Guidelines and man-
ufacturer's statement - Electromagnetic immunity
106 / 110
Interference im-
munity tests
IEC 60601 test levels
Compliance level
Electromagnetic en-
vironment -
Guidelines
Electrostatic discharge
(ESD) according to IEC
61000-4-2
± 6 kV contact discharge
± 8 kV air discharge
± 2/4/6 kV contact dis-
charge ± 2/4/8 kV air
discharge
Floors should be made of
wood or concrete or be
fitted with ceramic tiles.
If the floor is made of
synthetic material, the
relative humidity must
be at least 30%.
Fast transient electrical
disturbances/ bursts ac-
cording to IEC 6100-4-4
± 2 kV for power lines ±
1 kV for input and output
lines
± 2 kV for power lines
The quality of the supply
voltage should corres-
pond to that of a typical
business or hospital en-
vironment.
Surge voltage / surges
according to IEC
61000-4-5
± 1 kV push-pull voltage
± 2 kV common-mode
voltage
± 1 kV push-pull voltage
± 2 kV common-mode
voltage
The quality of the supply
voltage should corres-
pond to that of a typical
business or hospital en-
vironment.
Voltage interruptions,
short-term interruptions,
and fluctuations of the
supply voltage according
to IEC 61000-4-11
< 5 % V (> 95% inter-
ruption of the V T for ½
period
40% V (60% drop in V T
for 5 periods 70% V
(30% drop in V T for 25
periods < 5% V (> 95%
drop in V T for 5 s (250
periods)
< 5 % V (> 95% inter-
ruption of the V T for ½
period
40% V (60% drop in V T
for 5 periods 70% V
(30% drop in V T for 25
periods < 5% V (> 95%
drop in V T for 5 s (250
periods)
The quality of the supply
voltage should corres-
pond to that of a typical
business or hospital en-
vironment. If the user
requires the KaVoLUX
540 LED advanced func-
tions even in the case of
interruption in the power
supply, we recommend
the feeding of energy to
the KaVoLUX 540 LED
from an uninterruptible
power supply or from a
battery.
Magnetic field at a sup-
ply frequency (50/60 Hz)
according to IEC
61000-4-8
3 A/m
3 A/m
Magnetic fields at the
mains frequency should
correspond to the typical
values in a business and
hospital environment.
Note:
V T is the alternating mains voltage before the test level is used.
10.3 Guidelines and manufacturer's statement -
Electromagnetic immunity
The operating light KaVoLUX 540 LED is designed for operation in the environ-
ment defined below. The customer or user of the KaVoLUX 540 LED must en-
sure that it is operated in such an environment.