8 Safety check - Test instructions | 8.2 Ceiling mounting
Scan the light head of the treatment light KaVoLUX 540
LED with the probe
①
Fastening screw of the handle sup-
port when the gripping sleeve has
been removed
Equipment leakage current - Alternative measuring
method
Limit
< 10 mA (maximum value!)
ACP
Protection class 1
WARNING
Electrical power.
Death or injury from electric shock.
▶
Conduct test for leakage current in devices of Protection Class 1 only after
the protective earth test has been passed.
WARNING
Electrical power.
Death or injury from electric shock.
▶
Prior to connecting the treatment centre to the safety tester, disconnect the
treatment centre from the mains supply network.
Functional tests
The following conditions must be fulfilled in all function tests:
▪ The basic functions of the operating light must be guaranteed.
101 / 110