Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.11 Reprocessing KaVo Screen One / KaVo Screen HD
34 / 44
2.11 Reprocessing KaVo Screen One / KaVo Screen HD
Cleaning and disinfection
NOTICE
Non-colour-fast textiles.
Non-colour-fast textiles can cause permanent discoloration.
▶
Remove discolorations with water immediately.
NOTICE
Damage caused by spray disinfection.
Disinfectant gets into cracks and loosens the paint.
▶
Do not use spray disinfection.
▶
Only disinfect by wiping.
▶
Clean the surfaces with a soft cloth and water, and wipe-disinfect them with
an approved surface disinfectant.
Approved disinfectants:
▪ CaviCide / CaviWipes (Metrex)
▪ Microcide AF Liquid (Schülke & Mayr)
▪ FD 322 (Dürr Dental)
▪ Incidin Liquid (Ecolab)
2.12 Reprocessing of the service table 1568
Cleaning and disinfection
NOTICE
Non-colour-fast textiles.
Non-colour-fast textiles can cause permanent discoloration.
▶
Remove discolorations with water immediately.
NOTICE
Damage caused by spray disinfection.
Disinfectant gets into cracks and loosens the paint.
▶
Do not use spray disinfection.
▶
Only disinfect by wiping.