Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit
2.5.3 Reprocessing of sieve inserts and suction hoses
Note
Check the sieve inserts daily and after each surgery. The sieve inserts must
be replaced at the latest when the suction decreases (
Mat. no.
0.763.3423
).
Cleaning is not recommended for hygienic reasons.
▶
Remove the suction hoses from the holder.
ð
The suction motor starts automatically when the suction tubes are removed
from the holder.
▶
Hold the cloth under the sieve seat, and pull the sieve insert out of the
housing.
▶
If needed, insert a new sieve insert (
Mat. no. 07633423
) into the seat.
▶
Wipe the suction tubes with disinfectant.
2.6 Preparation of the patient unit
2.6.1 Preparation of the tumbler
▶
Reprocess as specified by the manufacturer.
2.6.2 Preparation of the tumbler holder and filler
NOTICE
Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or
"Bowl flush" keys.
▶
Make sure that the unit is turned off before removing the tumbler filler or
the spittoon bowl.
▶
Always check the correct seating of tumbler filler and spittoon bowl before
turning on the unit.