Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.10 Preparation of the amalgam separator
ter (limited rinsing). The water is aspirated from the Aquamat through the filter
housing to the separation system. The spittoon bowl rinsing is alternately
turned on and off.
The HYDROclean function prevents the coagulation and adhesion of proteins in-
side the drainage system and improves the transporting of fine particles.
Carrying out the HYDROclean function:
▪ After each treatment
Preparation
Requirement
Suction hoses remain hanging in the holder.
▶
Press the "HYDROclean" button.
ð
The HYDROclean function is in the preparatory stage.
Start the HYDROclean function
▶
Press the "HYDROclean" button.
ð
This starts-up the HYDROclean function.
The HYDROclean function is completed when the top LED goes dark.
The process can be discontinued at any time by pressing the "HYDROclean"
button.
2.10 Preparation of the amalgam separator
The following systems can be used as options:
▪ "Dürr" Amalgam separator
▪ Preparation for external wet suction for connection to a central amalgam
separator
▪ "Dürr" separation kit
Optimal care of the amalgam separation system includes the correct and regu-
lar use of the HYDROclean function and use of DEKASEPTOL gel.
See also:
2
2.9 Reprocessing and automatic hygiene functions of the suction system,
Page 32
Note
follow the instructions for use for the amalgam separator.
Replacing the amalgam container
▶
Turn off the treatment centre.
▶
Remove the cover.
▶
Replace the amalgam container according to the instructions for use of the
amalgam separator.