Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.6 Preparation of the patient unit
22 / 44
Cleaning and disinfection of the tumbler holder and
filler
Note
When assembling, make sure that the O-ring is present on the tumbler filler.
Note
Inspect the O-ring (
Mat. no. 0.200.6128
) for damage and exchange if ne-
cessary to prevent water or oxidation damage.
▶
Remove the tumbler filler
①
(
Mat. no. 0.764.8541
) by pulling upwards.
▶
Take off the tumbler holder
③
(
Mat. no. 07646153
).
▶
Clean the tumbler holder and tumbler filler under running water.
▶
Wet the O-ring
②
(
Mat. no. 02006128
) with silicone grease (
Mat. no.
10006403
).
NOTICE
Damage caused by treatment in a washer disinfector.
▶
Do not subject the instrument holder to washer disinfector treatment.
The tumbler filler can be subjected to thermodisinfection.
Sterilisation of the tumbler holder and filler
NOTICE
Damage caused by sterilisation.
▶
Do not sterilise the tumbler holder and filler.