Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting sys-
▶
Insert the triple-function or multifunctional handpieces without the cannula
into the clamping device of the germ reduction attachment. Make sure that
the "W" button (water) of the triple-function or multifunctional handpiece is
activated, and that no air is exiting.
▶
Place the instrument hoses of the dentist and assistant elements in the cor-
responding places of the germ reduction attachment.
▶
Support tubing with the enclosed holder.
Note
All water regulation mechanisms for the motor, instruments, and multiflex
couplings must be open.
Place the COMFORTbase on the rinsing attachment
Note
A seat adapter for the germ reduction attachment is available for the COM-
FORTbase.
▶
For the rinsing programme and intensive germ reduction, press the seat ad-
apter into one of the four rubber seats of the germ reduction attachment.
▶
Place the COMFORTbase in the seat adapter.
Carry out the rinsing programme
Set rinsing time
▶
Press the "decrease value" or "increase value" keys to select rinsing pro-
gram 01 "Programm vorbereiten: 01" or rinsing program 02 "Programm
vorbereiten: 02".
ð
The standard value for rinsing program 01 is 00:20 minutes per instrument.
ð
The standard value for rinsing program 02 is 2:00 minutes per instrument.
T
M
▶
Press the foot pedal to display the time for rinsing the instrument hoses and
the three-function or multifunctional handpiece.
With the foot pedal of the foot control being pressed down, the rinsing time for
the respective rinsing programme can be set in the range from 00:01 to 4:00
minutes using the "Decrease value" and "Increase value" buttons.
27 / 44