Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
Sterilisation of S-table handle and holder pad/anti-slip
mat
NOTICE
Damage caused by sterilisation.
▶
Do not sterilise the white standard handle.
▶
Sterilise the holder pad/ anti-slip mat.
The grey handle (
Mat. no. 1.002.4489
) can be sterilised (see symbol on the
handle).
2.4.2 Reprocessing of the instrument holder of the TM
table
Note
The handpiece holder (
Mat. no. 3.001.4229
) can be removed to facilitate
cleaning.
Cleaning and disinfection of the instrument holder
NOTICE
Damage caused by treatment in a washer disinfector.
▶
Do not treat the instrument holder in a washer disinfector.