Optisure Remote User’s Manual
Kahn Instruments
9
INSTALLATION
DANGER
Electric
Shock Risk
WARNING: Alarm leads MUST be potential free when wiring
to the connector block.
Figure 5
Alarm Output Connectors
For each output:
1.
Strip back the wire for the common (black) connection to the
COM
connector way for the
FAULT
alarm contact set, exposing approximately
0.25" (6mm) wire and clamp into the screw block
COM
terminal way.
Do
not overtighten the screw.
2.
Strip back the wire for the
N/O
(e.g. green) connection to the
N/O
connector way for the
FAULT
alarm contact set, exposing approximately
0.25" (6mm) wire and clamp into the screw block
N/O
terminal way.
Do
not overtighten the screw.
3.
Strip back the wire for the
N/C
(e.g. blue) connection to the
N/C
connector
way for the
FAULT
alarm contact set, exposing approximately 0.25" (6mm)
wire and clamp into the screw block
N/C
terminal way.
Do not overtighten
the screw.
4.
Repeat operations 1 to 3 for the
PROCESS
alarm contact set, using
appropriate colored wires.
5.
Locate the terminal block over the connector labelled
ALARMS
and push
the terminal block firmly into the connector.
Summary of Contents for Optisure Remote
Page 1: ...Optisure Remote Chilled Mirror Hygrometer User s Manual March 2018...
Page 2: ...Optisure Remote...
Page 7: ...Optisure Remote User s Manual Kahn Instruments vii NOTES...
Page 46: ...Optisure Remote User s Manual Kahn Instruments 39 APPENDIX B Appendix B...
Page 53: ...Optisure Remote User s Manual 46 March 2018 Optisure Series Variables Editor Screen...
Page 54: ...http www kahn com...