Optisure Remote User’s Manual
Kahn Instruments
17
OPERATION
3.1.4 Operational Status Display
ΔDP
Shows the total change in measured dew point over the time base of the trend graph
Mode
Shows current operation mode:
Measure, Standby, DCC, FAST, Max Cool, Data Hold
Next Mode
Counts down until the scheduled mode change, signifying either the start of a DCC or the
next measurement period
Process
Status of Process Alarm
For further information on alarm configuration see section 3.2.4.
Fault
Status of Fault Alarm
Indicates there is a fault with the optical system, clean the sensor by following the
instructions in section 4.8.
Sensor
Indicates whether the sensor has established a condensate formation, or if the system is
in a transient condition:
Heating, Cooling, Control
Logging
When shown, the Optisure is currently logging data to SD. See section 4.5.6 for
further information.
Table 4
Operational Status Display
3.1.5 Sensor Status Display
TEC Drive
Indicates whether the sensor is heating or cooling the mirror:
Also indicates the power level applied as a percentage of total
possible.
Optical Signal
Indicates the reflectivity of the mirror, and whether this is clean
or has a condensate formation.
The target is 100% signal level, which indicates the optimal film
thickness has been achieved. 0% indicates that the mirror is free
of condensate.
Table 5
Sensor Status Display
3.2
Setup Menus
3.2.1 DCC
Figure 14
DCC Screen
Summary of Contents for Optisure Remote
Page 1: ...Optisure Remote Chilled Mirror Hygrometer User s Manual March 2018...
Page 2: ...Optisure Remote...
Page 7: ...Optisure Remote User s Manual Kahn Instruments vii NOTES...
Page 46: ...Optisure Remote User s Manual Kahn Instruments 39 APPENDIX B Appendix B...
Page 53: ...Optisure Remote User s Manual 46 March 2018 Optisure Series Variables Editor Screen...
Page 54: ...http www kahn com...