6 Electrical connection
Slot
Option/
connec-
tion variant
Wire
(color)
a
Potential
Terminal
Symbol
DC
±5 V
Tempera-
ture input
Analysis
input
pH/redox
I
II
IN 7
IN 8
IN 9
IN 10
ISFET-pH
sensor
A (blue)
DC +5 V
1
B (black)
GND
with jump-
er to F
2
C (green)
DC -5 V
3
D (white/
black)
Ion-sensi-
tive gate
1
1
E
Bypass
3
4
5
5
F (yellow)
Reference
6
6
G (white)
Compensa-
tion ther-
mometer in
3-wire cir-
cuit
Connec-
tion
b
H (red)
I (red/black)
The RTD temperature probe is used to provide a temperature-com-
pensated pH-value measurement, and can be connected to a tem-
perature input or universal input.
c
a
The conductor colors listed refer to JUMO ISFET-pH sensors. The orange-colored conductor is not con-
nected.
b
The connection diagram for the selected analog input concerned must be observed when connecting the
temperature probe.
A
G
H
I
B
C
D
E
F
ϑ
G
H
I