8 Technical data
8.4
Electrical data
8.5
General Information
Voltage supply
In IO-Link operation
DC 18 to 32 V
In switch operation
DC 9.6 to 32 V
Nominal voltage
DC 24 V
The transmitter is designed only for operation in SELV and PELV circuits in ac-
cordance with DIN EN 50178.
Current consumption
In idle mode
≤ 12 mA (at nominal voltage)
In IO-Link operation
≤ 20 mA (at nominal voltage)
In switch operation
≤ 200 mA (at nominal voltage and with 2 switching outputs)
Electrical safety
Protection rating III according to DIN EN 61140
Intended use
Temperature measurement in industrial plants
Requirement
The transmitter must be equipped with an electrical circuit that meets the re-
quirements of EN 61010-1 with regard to "Limited-energy circuits".
Electrical connection
(input side)
Machine connector M12 × 1, 4-pole, according to DIN EN 61076-2-101
(socket version with union nut)
Electrical connection
(output side)
Machine connector M12 × 1, 4-pole, according to DIN EN 61076-2-101
(pin version)
Housing
Stainless steel
Protection type with mating connec-
tor
IP66, IP67, and IP69 according to DIN EN 60529
Installation position
Any
Weight
JUMO dTRANS T09 AS, type 707090 = ca. 35 g
JUMO dTRANS T09 DS, type 707091 = ca. 43 g
Potential equalization
Functional bonding conductor FB
a
a
The temperature probe must be connected to the plant's potential equalization system via the process connection.
Suitable shielded lines must also be used to ensure continuous shielding.
Summary of Contents for 00725428