8.29 Command SDS – Request status after executing shading correction command
This command returns the status of the shading correction function, with the following
parameters:
0=Shading correction not completed yet
1=Successful
3=Error1 – Timeout occurred
8.30 Command TG – Trigger Origin
Selects whether an external signal or an internal clock generator is used as a trigger
source.
Settings:
0=Internal clock generator
1=External signal
Associated commands:
TI
TP
8.31 Command TI – Trigger input
Selects whether the External Trigger input signal is taken from the Camera Link connector,
or from the 12-pin Hirose connector.
Settings:
0=Camera Link connector
1=12-pin Hirose connector
Important Note
The 12-pin Hirose connector trigger input can be
terminated for 75 ohm. This is selected by a rear
panel DIP-switch (SW1).
Fig.15 DIP switch
8.32 Command TP – Trigger polarity
Settings:
0=Active Low (factory default)
1=Active High
8.33 Command TR – Trigger Mode
Selects the trigger mode of the camera. Depending on the mode used, it allows the scan
rate to either be programmed by an internal timing generator or by and external trigger
pulse. See chapter 6 for details on the operation modes.
Settings:
0=No-Shutter mode
1=Shutter-Select mode
2=Pulse Width Control (PWC) mode
Associated functions:
Command TG (trigger origin)
Command TI (trigger input)
Command TP (trigger polarity)
1
2
Serial Comm. HR ⇔ CL
75Ω
ON ⇔ OFF
Factory default settings