Name
Description
Address
Range: 0 – 127
Baud rate
Select the baud rate from the following options: 5k/10k/20k/
40k/50k/80k/100k/125k/200k/250k/400k/500k/600k/800k/
1000k
Protocol
•
DeviceNet: Use standard can protocol.
•
BMS: Use the protocol for docking with the BMS
system.
The operation steps are as follows.
1. Press the composite keys
[Shift]+[P-set]
(System) on the front panel to en-
ter the system menu.
2. Rotate the knob or press the Up/Down key to select
I/O
and press
[Enter]
.
3. Press the Left/Right key to select
CAN
and press
[Enter]
.
4. Set the baud rate, address and other parameters, press
[Enter]
.
CAN Troubleshooting
If you meet some problems when communicating with PC by CAN interface,
please check the following items:
•
PC and the instrument must have the same baud rate.
•
Ensure you have used the correct communication cable (CAN_H, CAN_L).
Please pay attention that some cable may not have a correct internal wiring
even it is with an appropriate plug.
•
The interface cable is correctly connected (CAN_H to CAN_H, CAN_L to
CAN_L).
•
If the communication signal is poor or unstable, it is recommended to con-
nect a 120
Ω
terminating resistance.
–
The connection diagram of a single device is as below.
–
The connection diagram of multiple devices is as below.
Copyright © Itech Electronic Co., Ltd.
43
Summary of Contents for IT6000B Series
Page 130: ...Power Supply Function E 02 The waveform is as follows Copyright Itech Electronic Co Ltd 118...
Page 132: ...Power Supply Function E 06A The waveform is as follows Copyright Itech Electronic Co Ltd 120...
Page 136: ...Power Supply Function E 16 The waveform is as follows Copyright Itech Electronic Co Ltd 124...
Page 137: ...Power Supply Function E 18 The waveform is as follows Copyright Itech Electronic Co Ltd 125...
Page 150: ...Power Supply Function Copyright Itech Electronic Co Ltd 138...