Input
Pin 3 receives the level signal from the
outside.
Output
Pin 3 sends the digital signal (1, 0, PWM)
to the outside.
True
By default (
Not-Invert
), the out-
put digital signal is 1 (i.e. low lev-
el), and in the case of
Invert
, the
output is high level.
False
By default (
Not-Invert
), the out-
put digital signal is 0 (i.e. high
level), and in the case of
Invert
,
the output is low level.
PWM
Digital signal of PWM.
PWM Freq
Frequency
PWM Duty
Duty cycle
How to Use
When pin 3 is set to default
Off-Status
function, pin 3 will trigger level output
based on the
[On/Off]
state of the instrument. When it is set to (
Not-Invert
), the
[On/Off]
is turned off, and pin 3 outputs high level; the
[On/Off]
is turned on, and
pin 3 outputs low level.
When pin 3 is set to
Invert
, the output level is completely opposite.
1. Refer to the figure below to connect pin 3 to the external oscilloscope.