Press
[On/Off]
on the instrument’s front panel, the instrument’s output is
changed from On to Off. At this point, the pulse signal output from pin 7 can
be detected.
•
When the instrument’s
[On/Off]
is in Off state, input pulse signal to pin 7,
and it will not impact the
[On/Off]
state.
•
When the instrument’s
[On/Off]
is in On state, input pulse signal to pin 7,
and
[On/Off]
will be changed from On to Off.
Taking two instruments as an example, the operation is as below:
1. Referring to the figure below, connect pin 7 of two instruments to the exter-
nal oscilloscope.
2. Set pin 7’s function of two instruments to
Not-Invert
and
Sync-Off
.
3. Confirm that the
[On/Off]
of both instruments is in On state.
4. Press
[On/Off]
on the front panel of instrument A to turn off the output
function.
At this time, check the oscilloscope. The instrument A’s pin 7 outputs pulse
signal and the instrument B’s output function is synchronously turned off.
6.12 Source Analogue Function (Ext-Program)
(Optional)
This menu item is used to control whether the external analog function is turned
on. This function is not standard with the instrument and is optional for users. If
the user does not select this function, this menu item will not be displayed.
Copyright © Itech Electronic Co., Ltd.
205
Summary of Contents for IT6000B Series
Page 130: ...Power Supply Function E 02 The waveform is as follows Copyright Itech Electronic Co Ltd 118...
Page 132: ...Power Supply Function E 06A The waveform is as follows Copyright Itech Electronic Co Ltd 120...
Page 136: ...Power Supply Function E 16 The waveform is as follows Copyright Itech Electronic Co Ltd 124...
Page 137: ...Power Supply Function E 18 The waveform is as follows Copyright Itech Electronic Co Ltd 125...
Page 150: ...Power Supply Function Copyright Itech Electronic Co Ltd 138...