WatchPAT™200 System
35
Operation Manual
2.
Gently open the battery connector by disconnecting the 2 parts (you will need to
remove the transparent tape that secures the battery connector closed).
3.
Remove the battery.
4.
Insert the new battery into the battery compartment.
5.
Insert the
3 pin connector into the corresponding battery connector (one pin is
longer so it may properly be inserted in only one direction). Secure the battery
connector closed with a small piece of transparent tape.
6.
Close the battery compartment cover and secure back the screw.
Figure 24
– Replacing the Battery
Note
Please refer to the Service Manual or the support section of the Itamar
Medical web site for preventive maintenance recommendations.
6.11 Setting the Time and Date of the Watch
PAT™ Device
The WatchPAT™ device Time and Date can be set through the zzzPAT application.
Refer to the zzzPAT Software Manual for preparation of the WatchPAT™ device for a
new study.
Battery compartment cover
Phillips screw
Battery
Battery Connector - Closed
3 pin Connector