![Itamar Medical WatchPAT 200 Unified Operation Manual Download Page 66](http://html1.mh-extra.com/html/itamar-medical/watchpat-200-unified/watchpat-200-unified_operation-manual_2094893066.webp)
WatchPAT™200 System
58
Operation Manual
t
Action
Comment
E
xp
lai
n
t
o
S
u
b
je
ct
6.
Explain to
Subject
The subject may choose to perform
the night study during any night
during the next week.
It’s OK to shower with the bracelet.
Instruct subject to turn on the
WatchPAT™ device only after it is
connected to the bracelet.
THE BRACELET IS NOT TO BE
REMOVED UNTIL THE NIGHT
STUDY IS COMPLETED.
Removing the bracelet before or
during the night will cause the device
to shut down.
In the morning, instruct to cut the
bracelet along the dotted line, using
small scissors and to put it in the
WatchPAT™ device case along with
all the other parts (DO NOT
THROW BRACELET AWAY).
Do not try to connect ANY other
device to the bracelet.