WatchPAT™200 System
20
Operation Manual
CHARGER 22:51
Bat=3.12 V
Charging…
The display will show
“CHARGER”
if you are charging with the AC
adapter or
“PC HOST”
if you are charging with a computer.
The current battery voltage is shown.
Charge the battery the first time for approximately three hours.
Thereafter recharging takes approximately 1-1.5 hours.
5.
When charging is complete, the LCD will stop blinking and the following
screen will be displayed:
CHARGER 22:51
Bat=4.2 V
Charging complete
6.
Disconnect the AC adapter or communication cable. The WatchPAT™ device
will switch off in 30 seconds.
7.
Reseat the WatchPAT™ device in the wrist strap by gently sliding it back in
until a click is heard.
8.
Check that the oximetry module connector is properly connected to the
WatchPAT™.
Should a charging error arise the LCD will blink rapidly and the following screen will be
displayed.
CHARGER 22:51
Bat=4.2 V
Charger fault
3.2 Preparing the Oximetry Sensor
Use Nonin 8000JFW Flexi wrap pad and Nonin oximeter as supplied.