Table 2 - from IEC 60601-1-2:2007
Guidance and manufacturer's declaration
– electromagnetic immunity – WP200
The
WP200
is intended for use in the electromagnetic environment specified below; The customer or the user of the
WP200
should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2
Test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Electrical
fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines
2 kV for power
supply lines
ot Applicable
Mains power quality should be that of a typical public
low-voltage power supply network that supplies
buildings used for domestic purposes, commercial or
hospital, clinic environment.
Surge
IEC 61000-4-5
1 kV differential
mode
2 kV common mode
1 kV differential
mode
2 kV common mode
Mains power quality should be that of a typical public
low-voltage power supply network that supplies
buildings used for domestic purposes, commercial or
hospital, clinic environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
.
<5 %UT
(>95 %dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5 cycles
<5 %UT
70 %UT
(30 %dip in UT)
for 25 cycles
<5 %UT
<5 %UT
(>95 %dip in UT)
for 5 s
<5 %UT
(>95 %dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5 cycles
<5 %UT
70 %UT
(30 %dip in UT)
for 25 cycles
<5 %UT
<5 %UT
(>95 %dip in UT)
for 5 s
Mains power quality should be that of a typical public
low-voltage power supply network that supplies
buildings used for domestic purposes, commercial or
hospital, clinical environment.
If the user of the
WP200
requires continued
operation during power mains interruptions; it is
recommended that
WP200
be powered from a
separate power supply (UPS, etc.).
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical public low-voltage power
supply network that supplies buildings used for
domestic purposes, commercial or hospital, clinic
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.