WatchPAT™200 System
45
Operation Manual
Pull the paper tabs all the way off the body position
sensor while attaching it on the skin.
3.
Wearing the Wrist Strap
Should be comfortable, not too tight.
4.
Attaching the WatchPAT™ Device
Make sure the WatchPAT™ device is properly mounted
on the wrist strap. If it is loose, gently slide it in until you
hear a click.
5.
Attaching the Oximetry Sensor
Before PAT
®
probe attachment
Demonstrate proper placement of the finger on the Flexi
wrap pad – note the position of the fold line, and that the
two black square protrusions are opposite each other.
Folding the flaps of the Flexi wrap pad to secure sensor
properly.
Make sure it is not too tight
6.
Attaching the PAT
®
Probe
Insert finger all of the way into the probe
Press tip of probe against a hard surface (WatchPAT™
device case, table, etc.) while removing tabs in order to
keep the finger from moving inside the probe
Tab(s) must be fully removed
The probe is limited to a SINGLE USE. Do not remove
probe during the night.
8.4 Switching on the Watch
PAT™ Device
Demonstrate switching on the WatchPAT™ device by pressing the round
center button
Push button firmly until the LCD display lights up
The oximetry sensor light will glow red during the entire test
8.5 Removing the Watch
PAT™ Device
Demonstrate how to remove the WatchPAT™ device and place it back in
the carrying case.
The oximetry sensor light will keep glowing red.