WatchPAT™200 System
16
Operation Manual
At the completion of the device test, a
TEST FAILED
indicates a problem that
should be taken care of before the device is released for a night study.
DEVICE TEST 22:50
ID=111-11-1111
oxi=mod missing
pat=missing
<-Back More->
TEST FAILED 2:54
The following are the possible error, warning or information messages:
File error: not loaded, missing – the study file was not loaded or somehow
the file was deleted
File error: used x/3 x=1..3 – only when multi-night option is selected
Battery error: low – needs charging
Probe error: used, missing, bad – connect an unused probe
Oximetry error: module missing - connect oximetry module
Hardware (H/W) error: error code - contact customer support
SBP (Snore and Body Position sensor) warning: sensor missing – does not
affect PASSED status
RTC (Real Time Clock) warning: faulty – indicates problem with internal
clock but does not affect PASSED status
Bracelet error: missing – the study file was chosen with the bracelet option
but the bracelet is not connected during the device test
Information messages:
o
multi-night=on - when a multi night study is required
o
bracelet=on - when a study with tamper-proof patient
identification bracelet is required
More->
indicates that there are more error/warning messages and will be displayed if the
Right (
->
) button is pressed.
<-Back
will move to the previous screen if the Left (
<-
) button is pressed.
2.4.3 Oximetry Test
For the oximetry test make sure the sensor is attached to the finger. At the end of the
test the saturation and/or any error message will be displayed: