WatchPAT™200 System
47
Operation Manual
9
TROUBLESHOOTING GUIDE
9.1 Operator Error Messages
If an error message is displayed while performing the self-diagnostic tests, take the
actions specified below. If the problem persists contact Itamar or an authorized
representative.
Table 1
– Operator Troubleshooting
Error
Possible Reason
Action
File error
Not loaded
Study not initialized for new
patient
Connect device to PC and perform New
Study in zzzPAT
Battery error % full
Battery defective or uncharged
Charge battery or replace
Probe error
Used
Probe previously used
Replace probe
Missing
Probe absent
Attach probe
Bad
Probe is defective
Replace probe
Oximetry sensor error
No sensor
Sensor absent
Replace sensor
Disconnected
Sensor disconnected
Connect sensor cable to port
No communication
Module not connected
Check cable connection
Hardware status error code
WatchPAT™ device defective
Consult Itamar or authorized representative
SBP disconnected even if it
is connected
P200 or SBP sensor defective
Consult Itamar or authorized representative
RTC faulty
WatchPAT™ device defective
Consult Itamar or authorized representative
Short recording time
Patient removed the
WatchPAT™ device or probe
from hand prematurely
Explain proper use to patient
Insufficient battery charge
caused early termination of
recording
Recharge battery and try again
Damaged WatchPAT™ device
Contact your authorized sales
representative