Replacing electrical contacts
Electrical contacts should be inspected and replaced regularly as part of routine
maintenance. Overtime, electrical contacts can become depressed and worn out. To
ensure optimal life of the electrical contacts, wipe down the instrument and contacts
with a damp cloth after each use.
To replace an electrical contact, a #1 Phillips Screwdriver is required.
1.
Make sure the device power is OFF, then open the lid of the iBlot
™
2 Gel Transfer
Device.
2.
Hold the Lid with one hand, and remove the screw holding Contact 1. Discard
both the screw and the contact.
3.
Remove the screw holding Contact 2. Discard both the screw and the contact.
Appendix D
Maintenance
Replacing electrical contacts
D
iBlot
™
2 Dry Blotting System User Guide
51