![Invitrogen iBlot 2 User Manual Download Page 21](http://html1.mh-extra.com/html/invitrogen/iblot-2/iblot-2_user-manual_2085415021.webp)
A record of each run is kept by the iBlot
™
2 Gel Transfer Device, with voltage and
current being tracked at 1-second intervals. This information can be downloaded to a
USB storage device in “.csv” format.
1.
Insert a USB storage device into the USB port (Type A) of the iBlot
™
2 Gel
Transfer Device.
2.
Touch Write Logs to USB Device on the Options screen.
Note:
If you do not have a USB storage device in place, a prompt to insert a USB
storage device will appear. Touch OK to start the transfer after inserting the USB
storage device.
3.
Do not remove the USB storage device until the Options screen is displayed
again.
4.
The files are saved to folders titled “iBlot
™
2_Logs” and “iBlot
™
2_ErrorLogs” on
the USB storage device.
The App Notes button provides access to application notes for improving transfer
under various conditions when using the iBlot
™
2 Gel Transfer Device.