8 Troubleshooting
Troubleshooting
8.1 Troubleshooting the electrical system
Error
Display
(flashes at 0.5
sec. intervals
between two
codes)
Acoustic Signal
LED
(flashes
yellow)
Comment
Solution
Stoppage on
motor LEFT
Overload on
motor LEFT
E01
E0
10 beeps with 1
sec. ON and 1 sec.
OFF
When trying to
activate motor
again: 1 beep for
1 sec.
Left
only in Auto mode
(no display, acoustic signal
and flashing LED in manual
mode)
Check the cables to and from the
motor for the following:
- Cable undamaged.
- Cable correctly connected to motor.
- Cable connector correctly installed
and intact.
Stoppage on
motor RIGHT
Overload on
motor RIGHT
E02
E0
Right
Battery mode:
Power failure
OR
Emergency
stop activated
E03
E0
4 beeps with 1
sec. ON and 1 sec.
OFF
- 1st acoustic signal when
disconnected from power
supply OR the emergency
stop is activated. The sys-
tem switches to battery
mode and the wings will
stop moving and stay in
current position.
- 2nd acoustic signal after
20 min.
- 3rd acoustic signal after
40 min.
- After 60 min the system
switches to low battery
error (E04).
Connect to power supply
AND / OR
deactivate the emergency stop to
charge and check the battery (refer
to chapter Maintenance for further
information).
Low battery
E04
E0
4 beeps with 1
sec. ON and 1 sec.
OFF
- 1st acoustic signal when
the battery is detected to
be low while in battery
mode.
- Further acoustic signals
occur every 20 min until
the system turns off to
save battery power for
one CPR lowering.
System can
not be turned
on
Although the system is
connected to power supply
it can not be turned on.
Check if the emergency stop is
activated and deactivate if necessary.
Press the ON/OFF button to reset an ERROR and try again.
Contact your dealer or Invacare representative if the above does not solve your problems.