Electromagnetic compatibility (EMC)
10 Electromagnetic compatibility (EMC)
10.1 General EMC information
10.3 Electromagnetic Immunity
10.2 Electromagnetic emission
Medical Electrical Equipment needs to be installed and used according to the EMC information in this manual.
This product has been tested and found to comply with EMC limits specified by IEC/EN 60601-1-2 for Class B equipment.
Portable and mobile RF communications equipment can affect the operation of this product.
Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standard. To
determine if the emission from this product is causing the interference, run and stop running this product. If the interference with the
other device operation stops, then this product is causing the interference. In such rare cases, interference may be reduced or corrected
by the following:
• Reposition, relocate, or increase the separation between the devices.
Guidance and manufacturer ́s declaration
This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product should
assure that it is used in such an environment.
Guidance and manufacturer ́s declaration
This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product should
assure that it is used in such an environment.
Immunity test
Test / Compliance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8kV contact
± 2 kV, ± 4 kV, ± 8 kV, 15
kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic
material, the relative humidity should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power supply
lines; 100 kHz repetition
frequency
± 1 kV for input / output
lines; 100 kHz repetition
frequency
Mains power quality should be that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line to line
± 2 kV line to earth
Mains power quality should be that of a typical commercial or hospital environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 114
Group I
This product uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 114
Class B
This product is suitable for use in all establishments including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions IEC
61000-3-3
Complies