IVCNORBE221119
22
Invacare® Soft Tilt
* IPX6 classification means that the electrical system is protected
against water projected from any direction (not high pressure).
** IP66 classification means that the electrical system is protected
against high-pressure water jets projected from any direction
and fully protected against dust and other particulates, including
a vacuum seal.
9.5 Electrical system
Voltage supply: Uin 230 Voltage, AC, 50/60 Hz (AC = Alternating
current)
Maximum current input: Iin max.1,5 Ampere
Intermittent (periodic motor operation): Int = Max. 10 %, 2 min
ON / 18 min OFF
Insulation class: CLASS II
Type B Applied Part
Applied Part complying with the specified requirements for pro-
tection against electrical shock according to IEC60601-1.
(An applied parts is a part of the medical equipment which is
designed to come into physical contact with the patient or parts
that are likely to be brought into contact with the patient.)
Battery type: LP 12-0.8 (12 V 0.8AH) Sealed Lead-acid Battery
Constant voltage charge:
• Standby use: 13.5-13.8 V
• Cycle use: 14.4-15.0 V
• Initial current: Less than 0.24 A
Sound level: 58.5 dB (A)
Degree of protection: IPx6* or IP66**
The main module, actuators and hand control are protected
according to IPx6. The Control box is protected according to IP66
Summary of Contents for Soft Tilt